ECG & Pulse Doppler Systems For Cardiovascular Research on Mice
Electronics Contract R&D For Medical, Scientific, & Industrial Instrumentation

 
 
     
   
   
   
   
   
   
   
   
   

EXPERIMENTAL SETUP TO ACQUIRE DOPPLER BLOOD FLOW VELOCITY DATA FROM MICE


   

The overall experimental setup to acquire Doppler blood flow velocity data from mice is shown above. The mouse is anesthetized with Ketamine/Xylazine/Acepromazine cocktail, Nembutal cocktail, or other anesthetics depending on the desired type of study.

The mouse is placed in supine position with its four limbs taped to four electrode plates mounted on a board. These electrodes are connected via a cable to the ECG module in a signal processing system that will provide the mouse ECG signal output.

A 10 MHz pulsed Doppler probe is used with its tip placed just below the sternum and positioned to obtain a blood flow velocity signal from the aortic outlet or mitral inlet. In addition, blood flow velocities from peripheral arteries of mice can be obtained using a 20 MHz pulsed Doppler probe. These probes are plugged into analog pulsed Doppler signal generators (10 or 20 MHz) such as the “Baylor Multichannel High Frequency Pulsed Doppler Analog Mainframe”. Other equivalent systems such as those from Crystal Biotech / Data Sciences Inc. can also be used. These systems will generate in-phase and quadrature Doppler demodulated audio signals.

In the figure above, the Doppler signals are acquired, processed, and displayed as a real-time spectrogram along with the ECG signal using the DSPW. The software also allows the user to perform offline analysis of the data and to generate reports.

View Sample Reports from the DSPW